Lining | Electrode | ||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Fluid name | Chemical formula | concentration | Polyurethane rubber | ETFE | PFA | SUS316L | Titanium | Hastelloy C-276 | Tantalum | Platinum-iridium | Tungsten carbide | Nickel | Zirconium |
soldium chlorate | NaClO2 | 10 | ◎ | ◎ | △ 25℃ | ◎ | × | ◎ | ◎ | ||||
soldium chlorate | NaClO2 | 30 | ◎ | ◎ | △ 25℃ | ◎ | × | ◎ | ◎ | ||||
soldium chlorate | NaClO2 | 40 | ◎ | ◎ | △ 25℃ | × | × | ◎ | ◎ | ||||
sodium nitrate | NaNO2 | 60 | ◎ | ◎ | ◎ | ◎ | ◎ | ◎ | ◎ | ||||
arsenic, acid, arsenic | HAS2O3 | 10 | ◎ | ◎ | △ | △ | ◎ | ◎ | |||||
sulfurous acid | H2SO3 | 10 | ○ 35℃ | ◎ | ◎ | × | ○ 30℃ | △ | ◎ | ◎ | |||
sodium sulfite | Na2SO3 | 30 | ◎ | ◎ | ○75℃ △100℃ | ◎ | △ | ◎ | ◎ | ||||
aqueous ammonia | NH4OH | 28 | ○ 35℃ | ○ 45℃ | ○ 65℃ | ○25℃ △ | ○ 80℃ | ○25℃ △ | ◎ | × | △ | ||
Aquarius (sports drinks) | ◎ | △ | ◎ | ◎ | ◎ | ||||||||
drinking water | H2O | ○ 50℃ | ◎ | ◎ | ◎ | ◎ | ◎ | ◎ | ◎ | ◎ | ◎ | ◎ | |
Fluid name | Chemical formula | concentration | Polyurethane rubber | ETFE | PFA | SUS316L | Titanium | Hastelloy C-276 | Tantalum | Platinum-iridium | Tungsten carbide | Nickel | Zirconium |
zinc chloride | ZnCl2 | 20 | ◎ | ◎ | × | ◎ | ◎ | △ | ◎ | ||||
zinc chloride | ZnCl2 | 50 | ◎ | ◎ | × | ◎ | ◎ | × | ◎ | ||||
aluminum chloride | AlCl3 | 20 | ◎ | ◎ | × | × | ◎ | ◎ | ◎ | × | ◎ | ||
ammonium chloride | NH4Cl | 25 | ○ 50℃ | ◎ | ◎ | × | ◎ | ◎ | ◎ | ◎ | △ | ◎ | |
ammonium chloride | NH4Cl | 40 | ○ 50℃ | ◎ | ◎ | × | ◎ | × | ◎ | ◎ | △ | ◎ | |
cadmium chloride | CdCl2 | 10 | ◎ | ◎ | △ | △ 75℃ | △ 75℃ | ◎ | |||||
cadmium chloride | CdCl2 | 30 | ◎ | ◎ | △ | △ 25℃ | △ 25℃ | ◎ | |||||
potassium chloride | KCl | 10 | ○ 50℃ | ◎ | ◎ | × | ◎ | ○25℃ △ | ○25℃ △ | ◎ | △ | ◎ | |
potassium chloride | KCl | 40 | ○ 50℃ | ◎ | ◎ | × | ◎ | △ | △ | ◎ | △ 30% | ○ 30% | |
calcium chloride | CaCl2 | 70 | ○ 50℃ | ◎ | ◎ | × | ◎ | × | ◎ | ◎ | ○ 40% | ○ 60℃, 50% | |
Fluid name | Chemical formula | concentration | Polyurethane rubber | ETFE | PFA | SUS316L | Titanium | Hastelloy C-276 | Tantalum | Platinum-iridium | Tungsten carbide | Nickel | Zirconium |
tin chloride | SnCl2 | 10 | ◎ | ◎ | × | × | ◎ | ◎ | |||||
tin chloride | SnCl2 | 30 | ◎ | × | × | × | ◎ | ◎ | |||||
ferrous chloride | FeCl2 | 50 | ○ 50℃ | ◎ | ◎ | × | ◎ | △ | ◎ | ◎ | × | × | |
copper chloride | CuCl | 40 | ◎ | × | △ | △ 50℃ | ◎ | × | |||||
bichloride of mercury | HgCl2 | 40 | ◎ | ◎ | × | ◎ | △ 75℃ | ◎ | ◎ | ||||
tin chloride | SnCl4 | 20 | ◎ | × | △ | △ | ◎ | ◎ | |||||
tin chloride | SnCl4 | 50 | ◎ | × | × | △ | ◎ | ◎ | |||||
ferric chloride | FeCl3 | 10 | ◎ | ◎ | × | ◎ | ○ 50℃ | ◎ | ○ 25℃ | × | ◎ | ||
ferric chloride | FeCl3 | 30 | ◎ | ◎ | × | ◎ | △ | ◎ | ○ 25℃ | ◎ | |||
ferric chloride | FeCl3 | 50 | ◎ | ◎ | × | ◎ | × | ◎ | ○ 25℃ | ◎ | |||
Fluid name | Chemical formula | concentration | Polyurethane rubber | ETFE | PFA | SUS316L | Titanium | Hastelloy C-276 | Tantalum | Platinum-iridium | Tungsten carbide | Nickel | Zirconium |
copper chloride | CuCl2 | 40 | ○ 50℃ | ◎ | × | △ | △ 25℃ | ◎ | × | ||||
sodium chloride | NaCl | 25 | ○ 50℃ | ◎ | ◎ | × | ◎ | × | ◎ | ◎ | × | ◎ | |
nickel chloride | NiCl2 | 20 | ◎ | ◎ | × | ◎ | ◎ | ◎ | ○ 100% | ○ 100% | |||
barium chloride | BaCl2 | 25 | ○ 50℃ | ◎ | ◎ | × | ○ 50℃,20% | △ | ◎ | ◎ | △ | ○ 20% | |
magnesium chloride | MgCl2 | 40 | ○ 50℃ | ◎ | × | ◎ | ◎ | ◎ | ◎ | ||||
lithium chloride | LiCl | 60 | ○ 52℃ | ◎ | × | ◎ | ◎ | ◎ | ◎ | ||||
hydro- chloric acid | HCl | 10 | ◎ | ◎ | × | ○ 25℃ | ○ 25℃ | ○ ※1 | ○ 30℃ | × | ◎ | ||
hydro- chloric acid | HCl | 35 | ○ 70℃ | ◎ | × | × | △ 25℃ | ○ ※1 | ○ 30℃ | × | ○ 30% | ||
calcium chlorate | Ca(Col3)2 | 30 | ◎ | △ 1% | ○ 25℃ | △ | ◎ | ◎ | |||||
sodium chlorate | NaClO3 | 40 | ◎ | ◎ | × | ○ 25℃ | △ | ◎ | ◎ | ||||
yellow prussiate, potassium ferrocyanide | K4[Fe(CN)6] | 30 | ◎ | ◎ | △ | ◎ | △ | ◎ | ◎ | ◎ | |||
nitrohydro- chloric acid | 1HNO3+3HC | × | ◎ | ◎ | × | ○ 25℃ | × | ◎ | × | × | × |
Lining | Electrode | ||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Fluid name | Chemical formula | concentration | Polyurethane rubber | ETFE | PFA | SUS316L | Titanium | Hastelloy C-276 | Tantalum | Platinum-iridium | Tungsten carbide | Nickel | Zirconium |
sea water | ○ 50℃ | ◎ | ◎ | × | ◎ | △ | ◎ | ◎ | × | ||||
perchloric acid | HClO4 | 70 | ○ 40℃ | ◎ | × | ◎ | ◎ | × | |||||
hydrogen peroxide | H2O2 | 30 | × | ○ 45℃ | ○ 65℃ | ◎ | × | ○ 25℃ | ◎ | × | △ 60℃, 20% | ◎ | |
custic soda,sodium hydroxide | NaOH | 40 | ○ 35℃ | ◎ | ◎ | ※2 ○ | ◎ | △ | × | ◎ | ◎ | △ | |
custic soda,sodium hydroxide | NaOH | 50 | ○ 35℃ | ◎ | ◎ | ○ 80℃ | △ | △ | × | ◎ | ◎ | △ | |
potassium permanganate | KMnO4 | 30 | × | ○ 80℃ | ○ 90℃ | △ | △ | ◎ | ◎ | ||||
ammonium persulfate | (NH4)2S2O8 | 30 | × | ◎ | ◎ | ○50℃ △ | × | ◎ | ◎ | ||||
ammonium persulfate | (NH4)2S2O8 | 50 | × | ◎ | ◎ | ○ 50℃ | × | ◎ | ◎ | ||||
Fluid name | Chemical formula | concentration | Polyurethane rubber | ETFE | PFA | SUS316L | Titanium | Hastelloy C-276 | Tantalum | Platinum-iridium | Tungsten carbide | Nickel | Zirconium |
potassium persulfate | K2S2O8 | 10 | ◎ | ◎ | ◎ | ○ 75℃ | ○ 50℃ | ◎ | ◎ | ||||
formic acid | CH2O2 | 10 | × | ◎ | ◎ | × | × | ○50℃ △ | ◎ | ◎ | △ | ◎ | |
formic acid | CH2O2 | 50 | × | ◎ | ◎ | × | × | ○50℃ △ | ◎ | ◎ | △ | ◎ | |
milk | ◎ | ◎ | ◎ | ◎ | ◎ | ◎ | |||||||
citric acid | ○ 50℃ | ◎ | ◎ | ||||||||||
chromic acid | H2CrO4 | 50 | △ | ◎ | × | ◎ | △ | ○ ※1 | ◎ | × | ○ 30% | ||
sodium chromate | Na2CrO4 | 60 | △ | ◎ | ◎ | ◎ | ◎ | ○ ※1 | ◎ | ◎ | ◎ | ||
chloroacetic acid | C2H3ClO2 | 60 | ◎ | ◎ | × | △ | △ | ◎ | ◎ | ||||
chloro- sulfonic acid | HSO3l | 10 | × | ○ 20℃ | ○ 90℃ | × | × | △ 25℃ | ◎ | ◎ | |||
chloro- sulfonic acid | HSO3l | 45 | × | ○ 80℃ | ○ 90℃ | × | × | ○ 25℃ | ◎ | ◎ | |||
sodium silicate | Na2SiO3 | 10 | ◎ | ◎ | ◎ | ◎ | △ | ◎ | ◎ | ||||
hydrofluosi- licic acid | H2SiF6 | 35 | ◎ | ◎ | × | × | × | × | ◎ | ||||
antimony penta chlorate | SbCl5 | 50 | ◎ | ◎ | × | × | ◎ | ◎ | |||||
caffe | ◎ | ◎ | ◎ | ◎ | ◎ | ◎ |
Lining | Electrode | ||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Fluid name | Chemical formula | concentration | Polyurethane rubber | ETFE | PFA | SUS316L | Titanium | Hastelloy C-276 | Tantalum | Platinum-iridium | Tungsten carbide | Nickel | Zirconium |
acetic acid | C2H4O2 | 40 | × | ◎ | ◎ | ◎ | ◎ | ○ ※2 | ◎ | ◎ | × | ◎ | |
acetic acid | C2H4O2 | 75 | × | ◎ | ◎ | △ | ◎ | ○ ※2 | ◎ | ◎ | × | ◎ | |
sodium acetate | NaCH3CO2 | 100 | × | ◎ | ◎ | ○ 10% | ◎ | ◎ | ◎ | ◎ | ◎ | ◎ | |
lead acetate | Pb(CH3CO2)2 | 40 | × | ◎ | ◎ | △ | ○ 25℃ | × | ◎ | × | |||
antimony trichlorate | SbCl3 | 10 | ◎ | ◎ | × | × | ◎ | × | |||||
hypochlorous acid | HClO | 20 | ◎ | ◎ | × | × | ◎ | ◎ | × | ||||
calcium hypochlorite | Ca(ClO)2 | 20 | ○ 90℃ | ◎ | × | ◎ | △ | ◎ | ◎ | × | ◎ | ||
Fluid name | Chemical formula | concentration | Polyurethane rubber | ETFE | PFA | SUS316L | Titanium | Hastelloy C-276 | Tantalum | Platinum-iridium | Tungsten carbide | Nickel | Zirconium |
sodium hypochlorite | NaOCl | 15 | × | ○ 70℃ | ◎ | × | ◎ | ◎ | ◎ | ◎ | × | ○ 10℃ | |
sodium hypochlorite | NaOCl | 20 | × | ○ 90℃ | × | ◎ | ○ 50℃ | ◎ | ◎ | × | |||
copper cyanide | Cu(CN)2 | 10 | ○ 50℃ | ◎ | ◎ | ◎ | △ | ◎ | ◎ | ||||
sodium cyanide | NaCN | 10 | ◎ | ◎ | ◎ | ◎ | ◎ | ◎ | △ 40℃ | ◎ | |||
sodium cyanide | NaCN | 20 | ◎ | ◎ | × | ◎ | ◎ | ◎ | ◎ | ||||
hydrocyanic | CHN | ◎ | ○ 93℃ | × | × | × | ○ 93℃ | ○ | |||||
heavy sodium bisulfite | NaHSO3 | 40 | ◎ | ◎ | × | △ | △ | ◎ | ◎ | ||||
hydracid bromide | HBr | 48 | × | ○ 70℃ | ◎ | × | ◎ | × | ◎ | ◎ | |||
sodium bromide | NaBr | 50 | ◎ | ◎ | △ | × | △ | ◎ | ◎ | △ | |||
lithium bromide | LiBr | 60 | ◎ | × | ◎ | ◎ | |||||||
Fluid name | Chemical formula | concentration | Polyurethane rubber | ETFE | PFA | SUS316L | Titanium | Hastelloy C-276 | Tantalum | Platinum-iridium | Tungsten carbide | Nickel | Zirconium |
oxalic acid | C2H2O42H2O | 45 | ◎ | ◎ | × | × | △ | ◎ | ◎ | × | |||
heavy ammonium carbonate | NH4HCO3 | 50 | ◎ | ◎ | ◎ | △ | ◎ | ◎ | |||||
heavy potassium carbonate | KHCO3 | 30 | ◎ | ◎ | ◎ | ◎ | △ | ◎ | ◎ | ||||
heavy potassium carbonate | 50 | ◎ | ◎ | △ | ◎ | △ | ◎ | ◎ | |||||
heavy sodium carbonate | NaHCO3 | 20 | ◎ | ◎ | ◎ | ◎ | ◎ | ◎ | ◎ | ◎ | ◎ | ||
heavy ammonium fluoride | NH4HF2 | 5 | ◎ | ◎ | × | × | △ | × | ◎ | ||||
heavy sodium sulfide | NaHS | 40 | ◎ | ◎ | ◎ | ◎ | ◎ | ||||||
heavy sodium sulfite | NaHSO4 | 20 | ◎ | ◎ | × | ○ 25℃ | △ 50℃ | ◎ | ◎ | ||||
hitric acid | HNO3 | 10 | × | ○ 50℃ | ◎ | ◎ | ◎ | ○ 50℃ | ○ ※1 | ◎ | × | ||
hitric acid | HNO3 | 40 | × | ◎ | ○ 25℃ | ◎ | ○ 50℃ | ○ ※1 | ◎ | × | |||
Fluid name | Chemical formula | concentration | Polyurethane rubber | ETFE | PFA | SUS316L | Titanium | Hastelloy C-276 | Tantalum | Platinum-iridium | Tungsten carbide | Nickel | Zirconium |
hitric acid | HNO3 | 70 | × | ○ 90℃ | × | △ | × | ○ ※1 | ◎ | × | |||
alminum nitrate | Al(NO3)3 | 10 | ◎ | ◎ | ○ 75℃ | ◎ | △ 25℃ | ○ ※1 | ◎ | ||||
ammonium nitrate | NH4NO3 | 40 | × | ◎ | ◎ | ◎ | × | △ | ◎ | ◎ | ◎ | ||
potassium nitrate | KNO3 | 40 | ○ 50℃ | ◎ | ◎ | △ | ◎ | △ | ◎ | ◎ | △ | ||
calcium nitrate | Ca(NO3)2 | 10 | ○ | ◎ | ◎ | △ | ◎ | △ | ◎ | ◎ | △ | ||
silver nitrate | AgNO3 | 20 | ○ 50℃ | ◎ | ◎ | ○25℃ △ | ○ 50℃ | ○ 25℃ | ◎ | ◎ | × | ||
silver nitrate | AgNO3 | 50 | ○ 50℃ | ◎ | ◎ | ○25℃ △ | ○ 25℃ | ◎ | ◎ | × | |||
ferric nitrate | Fe(NO3)3 | 10 | ◎ | ◎ | △ | ○ 25℃ | ◎ | ◎ | |||||
ferric nitrate | Fe(NO3)3 | 50 | ◎ | ◎ | △ | △ 25℃ | ◎ | ◎ | |||||
copper nitrate | Cu(NO3)2 | 10 | ◎ | ◎ | ◎ | ○ 25℃ | △ 25℃ | ◎ | ◎ | ||||
Fluid name | Chemical formula | concentration | Polyurethane rubber | ETFE | PFA | SUS316L | Titanium | Hastelloy C-276 | Tantalum | Platinum-iridium | Tungsten carbide | Nickel | Zirconium |
copper nitrate | Cu(NO3)2 | 50 | ◎ | ◎ | △ | ○ 25℃ | △ 25℃ | ◎ | ◎ | ||||
sodium nitrate | NaNO3 | 30 | ◎ | ◎ | × | ◎ | ○ 25℃ | ◎ | ◎ | ◎ | |||
sodium nitrate | NaNO3 | 50 | ◎ | ◎ | × | ◎ | × | ◎ | ◎ | ◎ | |||
nickel nitrate | Ni(NO3)2 | 10 | ◎ | ◎ | △ | △ 25℃ | ◎ | ◎ | × | ||||
nickel nitrate | Ni(NO3)2 | 50 | ◎ | ◎ | △ 25℃ | × | ◎ | ◎ | × | ||||
magnesium nitrate | Mg(NO3)2 | 10 | ◎ | ◎ | △ | △ | △ | ◎ | ◎ | ||||
shochu with water | ◎ | ◎ | ◎ | ◎ | ◎ | ◎ | |||||||
soy sauce | ◎ | ◎ | × | ◎ | × | ◎ | ◎ | ||||||
alminum hydroxide | Al(OH)3 | 10 | ◎ | ◎ | △ | ◎ | △ | ◎ | ◎ | ◎ | ◎ | ||
ammonium hydroxide | NH4OH | 30 | ○ 50℃ | ○ 90℃ | ◎ | ○25℃ △ | ◎ | ○25℃ △ | × | ◎ | × | △ 30% | |
Fluid name | Chemical formula | concentration | Polyurethane rubber | ETFE | PFA | SUS316L | Titanium | Hastelloy C-276 | Tantalum | Platinum-iridium | Tungsten carbide | Nickel | Zirconium |
pottasium hydroxide | KOH | 10 | ○ 35℃ | ○ 80℃ | ○ 90℃ | × | ◎ | △ | ◎ | ◎ | ◎ | ◎ | |
pottasium hydroxide | KOH | 45 | ○ 35℃ | ○ 90℃ | × | × | △ | × | ◎ | ◎ | ◎ | ||
calcium hydroxide | Ca(OH)2 | 50 | ○ 50℃ | ◎ | △ | ◎ | △ | ◎ | ◎ | △ 10%, 100% | ○ 20% | ||
sodium hydroxide | NaOH | 40 | ○ 35℃ | ◎ | ◎ | ○ ※1 | ◎ ※1 | △ | × | ◎ | ◎ | △ | |
sodium hydroxide | NaOH | 50 | ○ 35℃ | ◎ | ◎ | ○ 80℃ | △ | △ | × | ◎ | ◎ | △ | |
barium hydroxide | Ba(OH)2 | 30 | ○ 50℃ | ◎ | ◎ | △ | ◎ | △ | × | ◎ | ◎ | ||
magnesium hydroxide | Mg(OH)2 | 10 | ○ | ◎ | ◎ | △ | △ 25℃ | × | ◎ | ||||
carbolic acid | C6H6O | 10 | ◎ | ◎ | △ | ◎ | ◎ | ◎ | |||||
potassium ferricyanide, red prussiate of potash | K3[Fe(CN)6] | 30 | ◎ | ◎ | △ | ◎ | △ | ◎ | × |
Lining | Electrode | ||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Fluid name | Chemical formula | concentration | Polyurethane rubber | ETFE | PFA | SUS316L | Titanium | Hastelloy C-276 | Tantalum | Platinum-iridium | Tungsten carbide | Nickel | Zirconium |
ammonium carbonate | (NH4)2CO3・H2O | 50 | ◎ | ◎ | △ | △ | ◎ | ◎ | |||||
potassium carbonate | K2CO3 | 20 | ◎ | ◎ | △ | ◎ | △ | × | ◎ | △ | |||
potassium carbonate | K2CO3 | 60 | ◎ | ◎ | △ | △ | △ | × | ◎ | △ | |||
calcium carbonate | CaCO3 | 10 | ◎ | ◎ | × | ◎ | × | ◎ | ◎ | △ | |||
sodium carbonate | Na2CO3 | 10 | ◎ | ◎ | × | ◎ | ◎ | ◎ | ◎ | ◎ | |||
sodium carbonate | Na2CO3 | 30 | ◎ | ◎ | × | ◎ | △ | ◎ | ◎ | ◎ | |||
barium carbonate | BaCO3 | 10 | ◎ | ◎ | △ | ○ 25℃ | × | ◎ | ◎ | ||||
magnesium carbonate | MgCO3 | 10 | ◎ | ◎ | △ | △ | △ | ◎ | ◎ | ||||
sodium thiosulfate | Na2S2O3 | 60 | ○ 50℃ | ◎ | ◎ | ◎ | △ | ◎ | ◎ |
Lining | Electrode | ||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Fluid name | Chemical formula | concentration | Polyurethane rubber | ETFE | PFA | SUS316L | Titanium | Hastelloy C-276 | Tantalum | Platinum-iridium | Tungsten carbide | Nickel | Zirconium |
chlorine dioxide | ClO2 | 15 | ◎ | ◎ | × | ◎ | △ | ◎ | ◎ | ||||
Japanese sake | ◎ | ◎ | ◎ | ◎ | ◎ | ◎ | |||||||
green tea | ◎ | ◎ | ◎ | ◎ | ◎ | ◎ | |||||||
uric acid | CH4N2O | 50 | ◎ | △ | ○ ※2 | ◎ | ◎ |
Lining | Electrode | ||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Fluid name | Chemical formula | concentration | Polyurethane rubber | ETFE | PFA | SUS316L | Titanium | Hastelloy C-276 | Tantalum | Platinum-iridium | Tungsten carbide | Nickel | Zirconium |
arsenic acid | H3AsO4 | 10 | ◎ | ◎ | △ | △ | ◎ | ◎ | △ | ||||
arsenic acid | H3AsO4 | 30 | ◎ | ◎ | △ | × | ◎ | ◎ | |||||
sodium hydrosulfite | NaHS | 10 | ◎ | ◎ | ◎ | ○ 25℃ | ◎ | ◎ | |||||
beer | ◎ | ◎ | ◎ | ◎ | ◎ | ◎ | |||||||
phenol | C6H6O | 10 | ◎ | ◎ | △ | ◎ | ◎ | ◎ | |||||
potassium ferrocyanide | K3[Fe(CN)6] | 30 | ◎ | △ | ◎ | △ | ◎ | × | △ | ||||
Fluid name | Chemical formula | concentration | Polyurethane rubber | ETFE | PFA | SUS316L | Titanium | Hastelloy C-276 | Tantalum | Platinum-iridium | Tungsten carbide | Nickel | Zirconium |
sodium ferricyanide | Na3[Fe(CN)6] | 10 | ◎ | △ 50℃ | △ 25℃ | ◎ | × | ||||||
potassium ferrocyanide | K4[Fe(CN)6] | 30 | ◎ | △ | ◎ | △ | ◎ | × | |||||
sodium ferrocyanide | Na[Fe(CN)6] | 10 | △ | △ | ◎ | × | |||||||
alminum fluoride | AlF3 | 20 | ◎ | × | × | △ 25℃ | ◎ | × | △ | ||||
ammonium fluoride | NH4F | 10 | ◎ | △ 25℃ | ○ 30℃ | ◎ | ◎ | × | × | ||||
ammonium fluoride | NH4F | 40 | ◎ | × | × | ◎ | ◎ | × | △ 20% | ||||
fluosilicate | H2SiF6 | 10 | ○ 90℃ | × | × | × | ◎ | × | |||||
fluoric acid | HF | 20 | × | ◎ | × | × | × | △ | ◎ | × | × | ||
fluoroboric acid | HBF4 | 20 | × | × | × | ◎ | × | ||||||
boracic acid | H3BO4 | 50 | ○ 90℃ | ◎ | × | ◎ | ◎ | ◎ | ◎ | △ 20% | |||
POCARI SWEAT (sports drink) | ◎ | △ | ◎ | ◎ | ◎ | ||||||||
formaldehyde | CH2O | 40 | ◎ | ◎ | ◎ | ◎ | △ | ◎ | ◎ |
Lining | Electrode | ||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Fluid name | Chemical formula | concentration | Polyurethane rubber | ETFE | PFA | SUS316L | Titanium | Hastelloy C-276 | Tantalum | Platinum-iridium | Tungsten carbide | Nickel | Zirconium |
chrome alum | KAl(SO4)2・12H2O | 10 | △ | △ | ◎ | ◎ | |||||||
water (drinking water) | H2O | ○ 50℃ | ◎ | ◎ | ◎ | ◎ | ◎ | ◎ | ◎ | ||||
water (seawater) | ○ 50℃ | ◎ | × | ◎ | △ | ◎ | ◎ | × | |||||
water (sewage) | ○ 50℃ | ◎ | ◎ | ◎ | ◎ | ◎ | ◎ | ◎ |
Lining | Electrode | ||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Fluid name | Chemical formula | concentration | Polyurethane rubber | ETFE | PFA | SUS316L | Titanium | Hastelloy C-276 | Tantalum | Platinum-iridium | Tungsten carbide | Nickel | Zirconium |
iodine | I | 10 | ◎ | × | × | △ | ◎ | ◎ |
Lining | Electrode | ||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Fluid name | Chemical formula | concentration | Polyurethane rubber | ETFE | PFA | SUS316L | Titanium | Hastelloy C-276 | Tantalum | Platinum-iridium | Tungsten carbide | Nickel | Zirconium |
ammonium sulfide | (NH4)2S | 40 | ◎ | △ | △ | ◎ | ◎ | ||||||
hydrogen sulfide | H2S | 10 | ◎ | ◎ | × | × | ◎ | ◎ | |||||
sodium sulfide | Na2S | 10 | ◎ | ◎ | ○75℃ △ | ◎ | △ | ◎ | ◎ | ||||
sodium sulfide | Na2S | 50 | ◎ | ◎ | × | × | △ | ◎ | ◎ | ||||
barium sulfide | BaS | 10 | ◎ | ◎ | △ 25℃ | ○ 25℃ | × | ○ 50℃ | ○ 50℃ | ○ 10℃ | |||
sulfuric acid | H2SO4 | 50 | ○ 35℃ | ◎ | ◎ | × | × | × | △ | ◎ | △ 10℃, 40% | △ | |
sulfuric acid | H2SO4 | 80 | × | ◎ | ◎ | × | × | × | △ | ◎ | × | △ 70% | |
sulfuric acid | H2SO4 | 96 | × | ◎ | ◎ | × | × | × | △ | ◎ | × | × | |
zinc sulfate | ZnSO4 | 40 | ◎ | ◎ | × | ◎ | △ | ◎ | ◎ | ||||
Fluid name | Chemical formula | concentration | Polyurethane rubber | ETFE | PFA | SUS316L | Titanium | Hastelloy C-276 | Tantalum | Platinum-iridium | Tungsten carbide | Nickel | Zirconium |
alminum sulfate | Al2(SO4)3・18H2O | 50 | ○ 50℃ | ○ 50℃ | ◎ | × | ◎ | ◎ | ◎ | ◎ | |||
potassium aluminium sulfate | KAl(SO4)3・18H2O | 10 | ◎ | △ | △ | ◎ | ◎ | ||||||
ammonium sulfate | (NH4)2SO4 | 20 | ○ 50℃ | ◎ | ◎ | △ | ◎ | △ | ◎ | ◎ | △ | ||
ammonium sulfate | (NH4)2SO4 | 40 | ○ | ◎ | ◎ | △ | × | △ | ◎ | ◎ | △ | ||
potassium sulfate | K2SO4 | 20 | ○ 50℃ | ◎ | ◎ | ◎ | ◎ | ◎ | ◎ | ◎ | △ | ◎ | |
calcium sulfate | Ca(SO4) | 10 | ○ 50℃ | ◎ | ◎ | ◎ | ○ 50℃ | × | ◎ | ◎ | |||
chrome sulfate | CrSO4 | 20 | × | ◎ | × | △ | ◎ | ◎ | |||||
ferrous sulfate | FeSO4 | 50 | ◎ | ◎ | × | ◎ | △ | ◎ | ◎ | ||||
ferric sulfate | Fe2(SO4)3 | 10 | ◎ | ◎ | ◎ | ○ 25℃ | ○50℃ △75℃ | ◎ | ◎ | × | ◎ | ||
ferric sulfate | Fe2(SO4)3 | 30 | ◎ | ◎ | × | ○ 25℃ | ○50℃ △75℃ | ◎ | ◎ | ◎ | |||
Fluid name | Chemical formula | concentration | Polyurethane rubber | ETFE | PFA | SUS316L | Titanium | Hastelloy C-276 | Tantalum | Platinum-iridium | Tungsten carbide | Nickel | Zirconium |
copper sulfate | CuSO4 | 30 | ◎ | ◎ | × | △ | ◎ | ◎ | ○ 50℃ | △ 10℃ | |||
sodium sulfate | Na2SO4 | 10 | ○ 50℃ | ◎ | ◎ | ◎ | △ | △ | ◎ | ◎ | △ | ◎ | |
sodium sulfate | Na2SO4 | 30 | ○ 50℃ | ◎ | ◎ | △ | △ | △ | ◎ | ◎ | △ | ◎ | |
nickel sulfate | NiSO4 | 10 | ○ 50℃ | ◎ | ◎ | △ | △ | ◎ | ◎ | ◎ 100% | ◎ 100% | ||
barium sulfate | BaSO4 | 10 | ○ 50℃ | ◎ | ◎ | △ | × | ◎ | ◎ | △ 100% | |||
magnesium sulfate | MgSO4 | 40 | ◎ | ◎ | ◎ | ◎ | × | ◎ | ◎ | △ 30% | |||
phosphoric acid | H3PO4 | 50 | ○ 50℃ | ◎ | ◎ | × | × | × | ○ ※3 | ◎ | × | △ | |
phosphoric acid | H3PO4 | 85 | ○ 50℃ | ◎ | ◎ | × | × | × | ○ ※3 | ◎ | × | × | |
ammonium phosphate | (NH4)3PO4 | 10 | ◎ | ◎ | × | ◎ | × | ◎ | ◎ | △ | ◎ | ||
sodium phosphate | Na2HPO4 | 50 | ◎ | ◎ | × | ◎ | ◎ | △ 25℃ | ◎ |
Lining | Electrode | ||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Fluid name | Chemical formula | concentration | Polyurethane rubber | ETFE | PFA | SUS316L | Titanium | Hastelloy C-276 | Tantalum | Platinum-iridium | Tungsten carbide | Nickel | Zirconium |
wine | ◎ | ◎ | ◎ | ◎ | ◎ | ◎ |