This is a microprocessor-based smart transmitter with excellent stability. It can measure the pressure of gases, liquids, and steam, as well as liquid levels.
- Safe instrumentation: IEC 61508-compliant and SIL2-certified
- Excellent long-term stability: zero point drift of only ±0.1 % over 10 years
- High-speed response within 100 ms for turbines and other applications that require fast response
Advanced Transmitter Gauge Pressure Transmitters (Model GTX_ _G) are used to measure the pressure of gases, liquids, and steam, as well as to measure tank pressure and fluid levels. They achieve high accuracy of ±0.04 % and long-term reliability (zero drift of only ±0.1 % over 10 years), greatly reducing calibration work after installation. With their very wide measurement range (GTX60G: 35 to 3500 kPa, a rangeability of 1:100), they help to cut costs by reducing the number of devices needed.
| Model | GTX60G | GTX71G | GTX82G |
|---|---|---|---|
| Measurement span | 17.5~3500 kPa | 0.7~14 MPa | 0.7~42 MPa |
| Setting range | -100~3500 kPa | -0.1~14 MPa | -0.1~42 MPa |
| Reference accuracy(Varies by material and measurement span) | ±0.04% | ±0.15% | ±0.15% |
| Fill fluid | Silicone oil (general purpose models), fluorine oil (oxygen and chlorine models) | ||
| Waterproof and dust-proof construction | IEC IP66/IP67 | ||
| Explosion-proof construction | TIIS flameproof (Exd IICT4) KOSHA flameproof FM explosion proof, intrinsic safe and Nonincendive ATEX/NEPSI/IECEx flameproof, intrinsic safe and Type n | ||
| Wetted material | SUS316 (diaphragm SUS316L), Alloy C-276, Tantalum | ||
| Power | 17.9~42V DC | ||
| Output signal | 4 to 20 mA DC (SFN communication) 4 to 20 mA DC (HART® communication) | ||
| Process connections | Rc 1/2 or 1/2 NPT internal thread | ||
| Safety certification | Safety Integrity Level (SIL) 2 certification | ||