Absolute Pressure Transmitters (Model JTA9_ _ _) are used to measure the absolute pressure of gases, liquids, and steam, as well as for monitoring tank pressure. Thanks to their very wide measurement range (Model JTA940S: 35 to 3500 kPa, a rangeability of 1:100), they help to cut costs by reducing the amount of equipment required.
Model | JTA922_ | JTA940_ |
---|---|---|
Measurement span | 4 to 104 kPa | 35 to 3500 kPa |
Setting range | 0 to 104 kPa abs | 0 to 3500 kPa abs |
Reference accuracy (Varies by material and measurement span) | ±0.15% | ±0.15% |
Fill fluid | Silicone oil (general purpose models), fluorine oil (oxygen and chlorine models) | |
Waterproof and dust-proof construction | IEC IP66/IP67 | |
Explosion-proof construction | TIIS explosion proof (Exd IICT4) | |
Wetted material | SUS316 (diaphragm SUS316L), Alloy C-276, Tantalum | |
Power | 17.9 to 45 V DC | |
Output signal | 4 to 20 mA DC (SFN communication) 4 to 20 mA DC (HART® communication) | |
Process connections | Rc 1/2 or 1/2 NPT internal thread |