Table for selecting wetted materials

 LiningElectrode
Fluid nameChemical formulaconcentrationPolyurethane rubberETFEPFASUS316LTitaniumHastelloy C-276TantalumPlatinum-iridiumTungsten carbideNickelZirconium
soldium chlorateNaClO210 ✓✓✓✓Conditional
25℃
✓✓×✓✓✓✓   
soldium chlorateNaClO230 ✓✓✓✓Conditional
25℃
✓✓×✓✓✓✓   
soldium chlorateNaClO240 ✓✓✓✓Conditional
25℃
××✓✓✓✓   
sodium nitrateNaNO260 ✓✓✓✓✓✓✓✓✓✓✓✓✓✓   
arsenic,
acid,
arsenic
HAS2O310 ✓✓✓✓Conditional Conditional✓✓✓✓   
sulfurous
acid
H2SO310✓
35℃
✓✓✓✓×✓
30℃
Conditional✓✓✓✓   
sodium sulfiteNa2SO330 ✓✓✓✓✓75℃
Conditional100℃
✓✓Conditional✓✓✓✓   
aqueous
ammonia
NH4OH28✓
35℃
✓
45℃
✓
65℃
✓25℃
Conditional
✓
80℃
✓25℃
Conditional
 ✓✓ ×Conditional
Aquarius
(sports drinks)
    ✓✓Conditional✓✓ ✓✓✓✓   
drinking waterH2O ✓
50℃
✓✓✓✓✓✓✓✓✓✓✓✓✓✓✓✓✓✓✓✓
Fluid nameChemical formulaconcentrationPolyurethane rubberETFEPFASUS316LTitaniumHastelloy C-276TantalumPlatinum-iridiumTungsten carbideNickelZirconium
zinc chlorideZnCl220 ✓✓✓✓×✓✓✓✓Conditional✓✓   
zinc chlorideZnCl250 ✓✓✓✓×✓✓✓✓×✓✓   
aluminum chlorideAlCl320 ✓✓✓✓××✓✓✓✓✓✓ ×✓✓
ammonium chlorideNH4Cl25✓
50℃
✓✓✓✓×✓✓✓✓✓✓✓✓ Conditional✓✓
ammonium chlorideNH4Cl40✓
50℃
✓✓✓✓×✓✓×✓✓✓✓ Conditional✓✓
cadmium chlorideCdCl210 ✓✓✓✓Conditional Conditional
75℃
Conditional
75℃
✓✓   
cadmium chlorideCdCl230 ✓✓✓✓Conditional Conditional
25℃
Conditional
25℃
✓✓   
potassium chlorideKCl10✓
50℃
✓✓✓✓×✓✓✓25℃
Conditional
✓25℃
Conditional
✓✓ Conditional✓✓
potassium chlorideKCl40✓
50℃
✓✓✓✓×✓✓ConditionalConditional✓✓ Conditional
30%
✓
30%
calcium
chloride
CaCl270✓
50℃
✓✓✓✓×✓✓×✓✓✓✓ ✓
40%
✓
60℃,
50%
Fluid nameChemical formulaconcentrationPolyurethane rubberETFEPFASUS316LTitaniumHastelloy C-276TantalumPlatinum-iridiumTungsten carbideNickelZirconium
tin
chloride
SnCl210  ✓✓✓✓××✓✓✓✓   
tin
chloride
SnCl230  ✓✓×××✓✓✓✓   
ferrous
chloride
FeCl250✓
50℃
✓✓✓✓×✓✓Conditional✓✓✓✓ ××
copper
chloride
CuCl40  ✓✓×ConditionalConditional
50℃
✓✓×   
bichloride
of mercury
HgCl240 ✓✓✓✓×✓✓Conditional
75℃
✓✓✓✓   
tin
chloride
SnCl420  ✓✓×ConditionalConditional✓✓✓✓   
tin
chloride
SnCl450  ✓✓××Conditional✓✓✓✓   
ferric
chloride
FeCl310 ✓✓✓✓×✓✓✓
50℃
✓✓✓
25℃
 ×✓✓
ferric
chloride
FeCl330 ✓✓✓✓×✓✓Conditional✓✓✓
25℃
  ✓✓
ferric
chloride
FeCl350 ✓✓✓✓×✓✓×✓✓✓
25℃
  ✓✓
Fluid nameChemical formulaconcentrationPolyurethane rubberETFEPFASUS316LTitaniumHastelloy C-276TantalumPlatinum-iridiumTungsten carbideNickelZirconium
copper
chloride
CuCl240✓
50℃
 ✓✓×ConditionalConditional
25℃
✓✓×   
sodium
chloride
NaCl 25✓
50℃
✓✓✓✓×✓✓×✓✓✓✓ ×✓✓
nickel
chloride
NiCl220 ✓✓✓✓×✓✓ ✓✓✓✓ ✓
100%
✓
100%
barium
chloride
BaCl225✓
50℃
✓✓✓✓×✓
50℃,20%
Conditional✓✓✓✓ Conditional✓
20%
magnesium
chloride
MgCl240✓
50℃
 ✓✓×✓✓✓✓✓✓✓✓   
lithium
chloride
LiCl60 ✓
52℃
✓✓×✓✓✓✓✓✓✓✓   
hydro-
chloric
acid
HCl10 ✓✓✓✓×✓
25℃
✓
25℃
✓
※1
✓
30℃
 ×✓✓
hydro-
chloric
acid
HCl35 ✓
70℃
✓✓××Conditional
25℃
✓
※1
✓
30℃
 ×✓
30%
calcium
chlorate
Ca(Col3)230  ✓✓Conditional
1%
✓
25℃
Conditional✓✓✓✓   
sodium
chlorate
NaClO340 ✓✓✓✓×✓
25℃
Conditional✓✓✓✓   
yellow prussiate, potassium ferrocyanideK4[Fe(CN)6]30 ✓✓✓✓ Conditional✓✓Conditional✓✓✓✓✓✓ 
nitrohydro-
chloric acid
1HNO3+3HC ×✓✓✓✓×✓
25℃
×✓✓× ××
 LiningElectrode
Fluid nameChemical formulaconcentrationPolyurethane rubberETFEPFASUS316LTitaniumHastelloy C-276TantalumPlatinum-iridiumTungsten carbideNickelZirconium
sea water  ✓
50℃
✓✓✓✓×✓✓Conditional✓✓✓✓×  
perchloric
acid
HClO470 ✓
40℃
✓✓×  ✓✓✓✓ × 
hydrogen
peroxide
H2O230×✓
45℃
✓
65℃
✓✓×✓
25℃
✓✓× Conditional
60℃,
20%
✓✓
custic
soda,sodium
hydroxide
NaOH40✓
35℃
✓✓✓✓※2
✓
✓✓Conditional×✓✓ ✓✓Conditional
custic
soda,sodium
hydroxide
NaOH50✓
35℃
✓✓✓✓✓
80℃
ConditionalConditional×✓✓ ✓✓Conditional
potassium
permanganate
KMnO430×✓
80℃
✓
90℃
Conditional Conditional✓✓✓✓   
ammonium
persulfate
(NH4)2S2O830×✓✓✓✓✓50℃
Conditional
 ×✓✓✓✓   
ammonium
persulfate
(NH4)2S2O850×✓✓✓✓✓
50℃
 ×✓✓✓✓   
Fluid nameChemical formulaconcentrationPolyurethane rubberETFEPFASUS316LTitaniumHastelloy C-276TantalumPlatinum-iridiumTungsten carbideNickelZirconium
potassium
persulfate
K2S2O810 ✓✓✓✓✓✓✓
75℃
✓
50℃
✓✓✓✓   
formic
acid
CH2O210×✓✓✓✓××✓50℃
Conditional
✓✓✓✓ Conditional✓✓
formic
acid
CH2O250×✓✓✓✓××✓50℃
Conditional
✓✓✓✓ Conditional✓✓
milk    ✓✓✓✓✓✓✓✓✓✓✓✓   
citric acid   ✓
50℃
✓✓✓✓       
chromic
acid
H2CrO450 Conditional✓✓×✓✓Conditional✓
※1
✓✓ ×✓
30%
sodium
chromate
Na2CrO460 Conditional✓✓✓✓✓✓✓✓✓
※1
✓✓ ✓✓✓✓
chloroacetic
acid 
C2H3ClO260 ✓✓✓✓×ConditionalConditional✓✓✓✓   
chloro-
sulfonic
acid
HSO3l10×✓
20℃
✓
90℃
××Conditional
25℃
✓✓✓✓   
chloro-
sulfonic
acid
HSO3l45×✓
80℃
✓
90℃
××✓
25℃
✓✓✓✓   
sodium
silicate
Na2SiO310 ✓✓✓✓✓✓✓✓Conditional✓✓✓✓   
hydrofluosi-
licic acid
H2SiF635 ✓✓✓✓××××✓✓   
antimony penta chlorateSbCl550 ✓✓✓✓× ×✓✓✓✓   
caffe    ✓✓✓✓✓✓✓✓✓✓✓✓   
 LiningElectrode
Fluid nameChemical formulaconcentrationPolyurethane rubberETFEPFASUS316LTitaniumHastelloy C-276TantalumPlatinum-iridiumTungsten carbideNickelZirconium
acetic acidC2H4O240×✓✓✓✓✓✓✓✓✓
※2
✓✓✓✓ ×✓✓
acetic acidC2H4O275×✓✓✓✓Conditional✓✓✓
※2
✓✓✓✓ ×✓✓
sodium
acetate
NaCH3CO2100×✓✓✓✓✓
10%
✓✓✓✓✓✓✓✓ ✓✓✓✓
lead
acetate
Pb(CH3CO2)240×✓✓✓✓Conditional✓
25℃
×✓✓×   
antimony trichlorateSbCl310 ✓✓✓✓× ×✓✓×   
hypochlorous
acid
HClO20 ✓✓✓✓××✓✓✓✓×   
calcium
hypochlorite
Ca(ClO)220 ✓
90℃
✓✓×✓✓Conditional✓✓✓✓ ×✓✓
Fluid nameChemical formulaconcentrationPolyurethane rubberETFEPFASUS316LTitaniumHastelloy C-276TantalumPlatinum-iridiumTungsten carbideNickelZirconium
sodium
hypochlorite
NaOCl15×✓
70℃
✓✓×✓✓✓✓✓✓✓✓ ×✓
10℃
sodium
hypochlorite
NaOCl20× ✓
90℃
×✓✓✓
50℃
✓✓✓✓ × 
copper
cyanide
Cu(CN)210✓
50℃
✓✓✓✓ ✓✓  Conditional✓✓ ✓✓    
sodium cyanideNaCN10 ✓✓✓✓✓✓✓✓ ✓✓✓✓ Conditional
40℃
✓✓
sodium cyanideNaCN20 ✓✓✓✓×✓✓ ✓✓✓✓  ✓✓
hydrocyanicCHN  ✓✓✓
93℃
×××✓
93℃
✓   
heavy sodium bisulfiteNaHSO340 ✓✓✓✓×ConditionalConditional✓✓✓✓   
hydracid bromideHBr48×✓
70℃
✓✓×✓✓×✓✓✓✓   
sodium bromide NaBr50 ✓✓✓✓Conditional×Conditional✓✓✓✓ Conditional 
lithium
bromide
LiBr60  ✓✓×  ✓✓✓✓   
Fluid nameChemical formulaconcentrationPolyurethane rubberETFEPFASUS316LTitaniumHastelloy C-276TantalumPlatinum-iridiumTungsten carbideNickelZirconium
oxalic acidC2H2O42H2O45 ✓✓✓✓××Conditional✓✓✓✓ × 
heavy ammonium carbonateNH4HCO350 ✓✓✓✓✓✓ Conditional✓✓✓✓   
heavy potassium carbonateKHCO330 ✓✓✓✓✓✓✓✓Conditional✓✓✓✓   
heavy potassium carbonate 50 ✓✓✓✓Conditional✓✓Conditional✓✓✓✓   
heavy sodium carbonateNaHCO320 ✓✓✓✓✓✓✓✓✓✓✓✓✓✓ ✓✓✓✓
heavy ammonium
fluoride
NH4HF25 ✓✓✓✓××Conditional×✓✓   
heavy sodium
sulfide
NaHS40 ✓✓✓✓✓✓  ✓✓✓✓   
heavy sodium
sulfite
NaHSO420 ✓✓✓✓×✓
25℃
Conditional
50℃
✓✓✓✓   
hitric acidHNO310×✓
50℃
✓✓✓✓✓✓✓
50℃
✓
※1
✓✓ × 
hitric acidHNO340× ✓✓✓
25℃
✓✓✓
50℃
✓
※1
✓✓ × 
Fluid nameChemical formulaconcentrationPolyurethane rubberETFEPFASUS316LTitaniumHastelloy C-276TantalumPlatinum-iridiumTungsten carbideNickelZirconium
hitric acidHNO370× ✓
90℃
×Conditional×✓
※1
✓✓ × 
alminum nitrateAl(NO3)310 ✓✓✓✓✓
75℃
✓✓Conditional
25℃
✓
※1
✓✓   
ammonium nitrateNH4NO340×✓✓✓✓✓✓×Conditional✓✓✓✓  ✓✓
potassium
nitrate
KNO340✓
50℃
✓✓✓✓Conditional✓✓Conditional✓✓✓✓ Conditional 
calcium nitrateCa(NO3)210✓✓✓✓✓Conditional✓✓Conditional✓✓✓✓ Conditional 
silver nitrateAgNO320✓
50℃
✓✓✓✓✓25℃
Conditional
✓
50℃
✓
25℃
✓✓✓✓ × 
silver nitrateAgNO350✓
50℃
✓✓✓✓✓25℃
Conditional
 ✓
25℃
✓✓✓✓ × 
ferric nitrateFe(NO3)310 ✓✓✓✓Conditional ✓
25℃
✓✓✓✓   
ferric nitrateFe(NO3)350 ✓✓✓✓Conditional Conditional
25℃
✓✓✓✓   
copper nitrateCu(NO3)210 ✓✓✓✓✓✓✓
25℃
Conditional
25℃
✓✓✓✓   
Fluid nameChemical formulaconcentrationPolyurethane rubberETFEPFASUS316LTitaniumHastelloy C-276TantalumPlatinum-iridiumTungsten carbideNickelZirconium
copper nitrateCu(NO3)250 ✓✓✓✓Conditional✓
25℃
Conditional
25℃
✓✓✓✓   
sodium nitrateNaNO330 ✓✓✓✓×✓✓✓
25℃
✓✓✓✓ ✓✓ 
sodium nitrateNaNO350 ✓✓✓✓×✓✓×✓✓✓✓ ✓✓ 
nickel nitrateNi(NO3)210 ✓✓✓✓Conditional Conditional
25℃
✓✓✓✓ × 
nickel nitrateNi(NO3)250 ✓✓✓✓Conditional
25℃
 ×✓✓✓✓ × 
magnesium
nitrate
Mg(NO3)210 ✓✓✓✓ConditionalConditionalConditional✓✓✓✓   
shochu with water    ✓✓✓✓✓✓✓✓✓✓✓✓   
soy sauce   ✓✓✓✓×✓✓×✓✓✓✓   
alminum hydroxideAl(OH)310 ✓✓✓✓Conditional✓✓Conditional✓✓✓✓ ✓✓✓✓
ammonium
hydroxide
NH4OH30✓
50℃
✓
90℃
✓✓✓25℃
Conditional
✓✓✓25℃
Conditional
×✓✓ ×
Conditional
30%
Fluid nameChemical formulaconcentrationPolyurethane rubberETFEPFASUS316LTitaniumHastelloy C-276TantalumPlatinum-iridiumTungsten carbideNickelZirconium
pottasium
hydroxide
KOH10✓
35℃
✓
80℃
✓
90℃
×✓✓Conditional✓✓✓✓ ✓✓✓✓
pottasium
hydroxide
KOH45✓
35℃
 ✓
90℃
××Conditional×✓✓ ✓✓✓✓
calcium hydroxideCa(OH)250✓
50℃
 ✓✓Conditional✓✓Conditional✓✓✓✓ Conditional
10%,
100%
✓
20%
sodium
hydroxide
NaOH40✓
35℃
✓✓✓✓✓
※1
✓✓
※1
Conditional×✓✓ ✓✓Conditional
sodium
hydroxide
NaOH50✓
35℃
✓✓✓✓✓
80℃
ConditionalConditional×✓✓ ✓✓Conditional
barium
hydroxide
Ba(OH)230✓
50℃
✓✓✓✓Conditional✓✓Conditional×✓✓ ✓✓ 
magnesium
hydroxide
Mg(OH)210✓✓✓✓✓Conditional Conditional
25℃
×✓✓   
carbolic acidC6H6O10 ✓✓✓✓Conditional ✓✓✓✓✓✓   
potassium
ferricyanide,
red prussiate
of potash
K3[Fe(CN)6]30 ✓✓✓✓Conditional✓✓Conditional✓✓×   
 LiningElectrode
Fluid nameChemical formulaconcentrationPolyurethane rubberETFEPFASUS316LTitaniumHastelloy C-276TantalumPlatinum-iridiumTungsten carbideNickelZirconium
ammonium
carbonate
(NH4)2CO3・H2O50 ✓✓✓✓Conditional Conditional✓✓✓✓   
potassium
carbonate
K2CO320 ✓✓✓✓Conditional✓✓Conditional×✓✓ Conditional 
potassium
carbonate
K2CO360 ✓✓✓✓ConditionalConditionalConditional×✓✓ Conditional 
calcium
carbonate
CaCO310 ✓✓✓✓×✓✓×✓✓✓✓ Conditional 
sodium carbonateNa2CO310 ✓✓✓✓×✓✓✓✓✓✓✓✓ ✓✓ 
sodium carbonateNa2CO330 ✓✓✓✓×✓✓Conditional✓✓✓✓ ✓✓ 
barium
carbonate
BaCO310 ✓✓✓✓Conditional✓
25℃
×✓✓✓✓   
magnesium
carbonate
MgCO310 ✓✓✓✓ConditionalConditionalConditional✓✓✓✓   
sodium thiosulfateNa2S2O360✓
50℃
✓✓✓✓✓✓ Conditional✓✓✓✓   
 LiningElectrode
Fluid nameChemical formulaconcentrationPolyurethane rubberETFEPFASUS316LTitaniumHastelloy C-276TantalumPlatinum-iridiumTungsten carbideNickelZirconium
chlorine
dioxide
ClO215 ✓✓✓✓×✓✓Conditional✓✓✓✓   
Japanese sake    ✓✓✓✓✓✓✓✓✓✓✓✓   
green tea    ✓✓✓✓✓✓✓✓✓✓✓✓   
uric acidCH4N2O50  ✓✓Conditional✓
※2
 ✓✓✓✓   
 LiningElectrode
Fluid nameChemical formulaconcentrationPolyurethane rubberETFEPFASUS316LTitaniumHastelloy C-276TantalumPlatinum-iridiumTungsten carbideNickelZirconium
arsenic acidH3AsO410 ✓✓✓✓Conditional Conditional✓✓✓✓ Conditional 
arsenic acidH3AsO430 ✓✓✓✓Conditional ×✓✓✓✓   
sodium hydrosulfiteNaHS10 ✓✓✓✓✓✓ ✓
25℃
✓✓✓✓   
beer    ✓✓✓✓✓✓✓✓✓✓✓✓   
phenolC6H6O10 ✓✓✓✓Conditional ✓✓✓✓✓✓   
potassium ferrocyanideK3[Fe(CN)6]30 ✓✓ Conditional✓✓Conditional✓✓× Conditional 
Fluid nameChemical formulaconcentrationPolyurethane rubberETFEPFASUS316LTitaniumHastelloy C-276TantalumPlatinum-iridiumTungsten carbideNickelZirconium
sodium ferricyanideNa3[Fe(CN)6]10  ✓✓Conditional
50℃
 Conditional
25℃
 ✓✓×  
potassium
ferrocyanide
K4[Fe(CN)6]30 ✓✓ Conditional✓✓Conditional✓✓×   
sodium ferrocyanideNa[Fe(CN)6]10    Conditional Conditional✓✓×  
alminum
fluoride
AlF320  ✓✓ ××Conditional
25℃
✓✓× Conditional
ammonium
fluoride
NH4F10  ✓✓ Conditional
25℃
✓
30℃
✓✓✓✓× ×
ammonium
fluoride
NH4F40  ✓✓ ××✓✓✓✓× Conditional
20%
fluosilicateH2SiF610  ✓
90℃
 ×××✓✓×  
fluoric acidHF20 ×✓✓×××Conditional✓✓× ×
fluoroboric acidHBF420    ×××✓✓×  
boracic acidH3BO450  ✓
90℃
✓✓×✓✓✓✓✓✓✓✓ Conditional
20%
POCARI SWEAT (sports drink)    ✓✓Conditional✓✓ ✓✓✓✓   
formaldehydeCH2O40  ✓✓✓✓✓✓✓✓Conditional✓✓✓✓  
 LiningElectrode
Fluid nameChemical formulaconcentrationPolyurethane rubberETFEPFASUS316LTitaniumHastelloy C-276TantalumPlatinum-iridiumTungsten carbideNickelZirconium
chrome alumKAl(SO4)2・12H2O10    Conditional Conditional✓✓✓✓  
water (drinking water)H2O  ✓
50℃
✓✓ ✓✓✓✓✓✓✓✓✓✓✓✓ 
water (seawater)   ✓
50℃
✓✓×✓✓Conditional✓✓✓✓×  
water (sewage)   ✓
50℃
✓✓ ✓✓✓✓✓✓✓✓✓✓✓✓ 
 LiningElectrode
Fluid nameChemical formulaconcentrationPolyurethane rubberETFEPFASUS316LTitaniumHastelloy C-276TantalumPlatinum-iridiumTungsten carbideNickelZirconium
iodineI10  ✓✓ ××Conditional✓✓✓✓  
 LiningElectrode
Fluid nameChemical formulaconcentrationPolyurethane rubberETFEPFASUS316LTitaniumHastelloy C-276TantalumPlatinum-iridiumTungsten carbideNickelZirconium
ammonium
sulfide
(NH4)2S40  ✓✓Conditional Conditional✓✓✓✓   
hydrogen
sulfide
H2S10 ✓✓✓✓× ×✓✓✓✓   
sodium
sulfide
Na2S10 ✓✓✓✓✓75℃
Conditional
✓✓Conditional✓✓✓✓   
sodium
sulfide
Na2S50 ✓✓✓✓××Conditional✓✓✓✓   
barium
sulfide
BaS10 ✓✓✓✓Conditional
25℃
✓
25℃
×✓
50℃
✓
50℃
  ✓
10℃
sulfuric acidH2SO450✓
35℃
✓✓✓✓×××Conditional✓✓ Conditional
10℃,
40%
Conditional
sulfuric acidH2SO480×✓✓✓✓×××Conditional✓✓ ×Conditional
70%
sulfuric acidH2SO496×✓✓✓✓×××Conditional✓✓ ××
zinc sulfateZnSO440 ✓✓✓✓×✓✓Conditional✓✓✓✓   
Fluid nameChemical formulaconcentrationPolyurethane rubberETFEPFASUS316LTitaniumHastelloy C-276TantalumPlatinum-iridiumTungsten carbideNickelZirconium
alminum sulfateAl2(SO4)3・18H2O50✓
50℃
✓
50℃
✓✓×✓✓✓✓✓✓✓✓   
potassium aluminium sulfateKAl(SO4)3・18H2O10  ✓✓Conditional Conditional✓✓✓✓   
ammonium
sulfate
(NH4)2SO420✓
50℃
✓✓✓✓Conditional✓✓Conditional✓✓✓✓ Conditional 
ammonium
sulfate
(NH4)2SO440✓✓✓✓✓Conditional×Conditional✓✓✓✓ Conditional 
potassium
sulfate
K2SO420✓
50℃
✓✓✓✓✓✓✓✓✓✓✓✓✓✓ Conditional✓✓
calcium
sulfate
Ca(SO4)10✓
50℃
✓✓✓✓✓✓✓
50℃
×✓✓✓✓   
chrome
sulfate
CrSO420× ✓✓× Conditional✓✓✓✓   
ferrous
sulfate
FeSO450 ✓✓✓✓×✓✓Conditional✓✓✓✓   
ferric
sulfate
Fe2(SO4)310 ✓✓✓✓✓✓✓
25℃
✓50℃
Conditional75℃
✓✓✓✓ ×✓✓
ferric
sulfate
Fe2(SO4)330 ✓✓✓✓×✓
25℃
✓50℃
Conditional75℃
✓✓✓✓  ✓✓
Fluid nameChemical formulaconcentrationPolyurethane rubberETFEPFASUS316LTitaniumHastelloy C-276TantalumPlatinum-iridiumTungsten carbideNickelZirconium
copper
sulfate
CuSO430 ✓✓✓✓×Conditional✓✓✓✓✓
50℃
 Conditional
10℃
 
sodium
sulfate
Na2SO410✓
50℃
✓✓✓✓✓✓ConditionalConditional✓✓✓✓ Conditional✓✓
sodium
sulfate
Na2SO430✓
50℃
✓✓✓✓ConditionalConditionalConditional✓✓✓✓ Conditional✓✓
nickel
sulfate 
NiSO410✓
50℃
✓✓✓✓Conditional Conditional✓✓✓✓ ✓✓
100%
✓✓
100%
barium
sulfate
BaSO410✓
50℃
✓✓✓✓Conditional ×✓✓✓✓ Conditional
100%
 
magnesium
sulfate
MgSO440 ✓✓✓✓✓✓✓✓×✓✓✓✓ Conditional
30%
 
phosphoric acidH3PO450✓
50℃
✓✓✓✓×××✓
※3
✓✓ ×Conditional
phosphoric acidH3PO485✓
50℃
✓✓✓✓×××✓
※3
✓✓ ××
ammonium
phosphate
(NH4)3PO410 ✓✓✓✓×✓✓×✓✓✓✓ Conditional✓✓
sodium
phosphate
Na2HPO450 ✓✓✓✓×✓✓✓✓Conditional
25℃
✓✓   
 LiningElectrode
Fluid nameChemical formulaconcentrationPolyurethane rubberETFEPFASUS316LTitaniumHastelloy C-276TantalumPlatinum-iridiumTungsten carbideNickelZirconium
wine    ✓✓✓✓✓✓✓✓✓✓✓✓   

Meaning of symbols in the table

The symbols in the selection table indicate the following.

SymbolContent
✓✓Usable (recommended material) a), 100 ° C or less at indicated concentration
✓There are restrictions on fluid conditions such as concentration and temperature, but it can be used up to the indicated temperature
ConditionalCorrosion resistance is lower than the above symbol  b), and it can only be used as a grounding ring material  c)
×Not available
No symbolNo data
※1, ※2, ※3Some applications require special attention, so please contact us separately.

a) The usable limit for electrode materials is a corrosion rate of 0.05 mm/year or less. Even if the corrosion rate is lower than the above, materials that may cause intergranular corrosion or stress corrosion cannot be used.
b) The usable limit for the grounding ring material is 0.5 mm/year or less.
c) Can be used up to the indicated temperature. If there is no indication, the maximum temperature is 100℃.